Nuclear magnetic resonance (NMR) is a physical phenomenon in which nuclei in a magnetic field absorb and re-emit electromagnetic radiation. This energy is at a specific resonance frequency which depends on the strength of the magnetic field and the magnetic properties of the isotope of the atoms; in practical applications, the frequency is similar to VHF and UHF television broadcasts (60-1000 MHz). NMR allows the observation of specific quantum mechanical magnetic properties of the atomic nucleus. Many scientific techniques exploit NMR phenomena to study molecular physics, crystals, and non-crystalline materials through nuclear magnetic resonance spectroscopy. NMR is also routinely used in advanced medical imaging techniques, such as in magnetic resonance imaging (MRI).

All isotopes that contain an odd number of protons and/or neutrons (see Isotope) have an intrinsic magnetic moment and angular momentum, in other words a nonzero spin, while all nuclides with even numbers of both have a total spin of zero. The most commonly studied nuclei are 1
H
and 13
C
, although nuclei from isotopes of many other elements (e.g. 2
H
, 6
Li
, 10
B
, 11
B
, 14
N
, 15
N
, 17
O
, 19
F
, 23
Na
, 29
Si
, 31
P
, 35
Cl
, 113
Cd
, 129
Xe
, 195
Pt
) have been studied by high-field NMR spectroscopy as well.

A key feature of NMR is that the resonance frequency of a particular substance is directly proportional to the strength of the applied magnetic field. It is this feature that is exploited in imaging techniques; if a sample is placed in a non-uniform magnetic field then the resonance frequencies of the sample's nuclei depend on where in the field they are located. Since the resolution of the imaging technique depends on the magnitude of magnetic field gradient, many efforts are made to develop increased field strength, often using superconductors. The effectiveness of NMR can also be improved using hyperpolarization, and/or using two-dimensional, three-dimensional and higher-dimensional multi-frequency techniques.

The principle of NMR usually involves two sequential steps:

  • The alignment (polarization) of the magnetic nuclear spins in an applied, constant magnetic field B0.
  • The perturbation of this alignment of the nuclear spins by employing an electro-magnetic, usually radio frequency (RF) pulse. The required perturbing frequency is dependent upon the static magnetic field (H0) and the nuclei of observation.

The two fields are usually chosen to be perpendicular to each other as this maximizes the NMR signal strength. The resulting response by the total magnetization (M) of the nuclear spins is the phenomenon that is exploited in NMR spectroscopy and magnetic resonance imaging. Both use intense applied magnetic fields (H0) in order to achieve dispersion and very high stability to deliver spectral resolution, the details of which are described by chemical shifts, the Zeeman effect, and Knight shifts (in metals).

NMR phenomena are also utilized in low-field NMR, NMR spectroscopy and MRI in the Earth's magnetic field (referred to as Earth's field NMR), and in several types of magnetometers.


Video Nuclear magnetic resonance



History

Nuclear magnetic resonance was first described and measured in molecular beams by Isidor Rabi in 1938, by extending the Stern-Gerlach experiment, and in 1944, Rabi was awarded the Nobel Prize in Physics for this work. In 1946, Felix Bloch and Edward Mills Purcell expanded the technique for use on liquids and solids, for which they shared the Nobel Prize in Physics in 1952.

Yevgeny Zavoisky likely observed nuclear magnetic resonance in 1941, well before Felix Bloch and Edward Mills Purcell, but dismissed the results as not reproducible.

Russell H. Varian filed the "Method and means for correlating nuclear properties of atoms and magnetic fields", U.S. Patent 2,561,490 on July 24, 1951. Varian Associates developed the first NMR unit called NMR HR-30 in 1952.

Purcell had worked on the development of radar during World War II at the Massachusetts Institute of Technology's Radiation Laboratory. His work during that project on the production and detection of radio frequency power and on the absorption of such RF power by matter laid the foundation for Rabi's discovery of NMR.

Rabi, Bloch, and Purcell observed that magnetic nuclei, like 1
H
and 31
P
, could absorb RF energy when placed in a magnetic field and when the RF was of a frequency specific to the identity of the nuclei. When this absorption occurs, the nucleus is described as being in resonance. Different atomic nuclei within a molecule resonate at different (radio) frequencies for the same magnetic field strength. The observation of such magnetic resonance frequencies of the nuclei present in a molecule allows any trained user to discover essential chemical and structural information about the molecule.

The development of NMR as a technique in analytical chemistry and biochemistry parallels the development of electromagnetic technology and advanced electronics and their introduction into civilian use.


Maps Nuclear magnetic resonance



Theory of nuclear magnetic resonance

Nuclear spin and magnets

All nucleons, that is neutrons and protons, composing any atomic nucleus, have the intrinsic quantum property of spin. The overall spin of the nucleus is determined by the spin quantum number S. If the numbers of both the protons and neutrons in a given nuclide are even then S = 0, i.e. there is no overall spin. Then, just as electrons pair up in atomic orbitals, so do even numbers of protons or even numbers of neutrons, (both of which are also spin 1 2 {\textstyle {\frac {1}{2}}} particles and hence fermions) pair up giving zero overall spin.

However, a proton and neutron will have lower energy when their spins are parallel, not anti-parallel. Parallel spin alignment does not infringe upon the Pauli Exclusion Principle. The lowering of energy for parallel spins has to do with the quark structure of these two nucleons. Therefore, the spin ground state for the deuteron (the deuterium nucleus, or the 2H isotope of hydrogen)--that has only a proton and a neutron--corresponds to a spin value of 1, not of zero. The single, isolated deuteron therefore exhibits an NMR absorption spectrum characteristic of a quadrupolar nucleus of spin 1, which in the "rigid" state at very low temperatures is a characteristic ('Pake') doublet, (not a singlet as for a single, isolated 1H, or any other isolated fermion or dipolar nucleus of spin 1/2). On the other hand, because of the Pauli Exclusion Principle, the tritium isotope of hydrogen must have a pair of anti-parallel spin neutrons (of total spin zero for the neutron-spin pair), plus a proton of spin 1/2. Therefore, the character of the tritium nucleus is again magnetic dipolar, not quadrupolar--like its non-radioactive deuteron cousin--and the tritium nucleus total spin value is again 1/2, just like for the simpler, abundant hydrogen isotope, 1H nucleus (the proton). The NMR absorption (radio) frequency for tritium is however slightly higher than that of 1H because the tritium nucleus has a slightly higher gyromagnetic ratio than 1H. In many other cases of non-radioactive nuclei, the overall spin is also non-zero. For example, the 27
Al
nucleus has an overall spin value S = 5/2.

A non-zero spin is thus always associated with a non-zero magnetic moment (?) via the relation ? = ?S, where ? is the gyromagnetic ratio. It is this magnetic moment that allows the observation of NMR absorption spectra caused by transitions between nuclear spin levels. Most nuclides (with some rare exceptions) that have both even numbers of protons and even numbers of neutrons, also have zero nuclear magnetic moments, and they also have zero magnetic dipole and quadrupole moments. Hence, such nuclides do not exhibit any NMR absorption spectra. Thus, 18
O
is an example of a nuclide that has no NMR absorption, whereas 13
C
, 31
P
, 35
Cl
and 37
Cl
are nuclides that do exhibit NMR absorption spectra. The last two nuclei are quadrupolar nuclei whereas the preceding two nuclei (13
C
and 31
P
) are dipolar ones.

Electron spin resonance (ESR) is a related technique in which transitions between electronic spin levels are detected rather than nuclear ones. The basic principles are similar but the instrumentation, data analysis, and detailed theory are significantly different. Moreover, there is a much smaller number of molecules and materials with unpaired electron spins that exhibit ESR (or electron paramagnetic resonance (EPR)) absorption than those that have NMR absorption spectra. ESR has much higher sensitivity than NMR does.

Values of spin angular momentum

The angular momentum associated with nuclear spin is quantized. This means both that the magnitude of angular momentum is quantized (i.e. S can only take on a restricted range of values), and also that the orientation of the associated angular momentum is quantized. The associated quantum number is known as the magnetic quantum number, m, and can take values from +S to -S, in integer steps. Hence for any given nucleus, there are a total of 2S + 1 angular momentum states.

The z-component of the angular momentum vector (S) is therefore Sz = m?, where ? is the reduced Planck constant. The z-component of the magnetic moment is simply:

? z = ? S z = ? m ? . {\displaystyle \mu _{z}=\gamma S_{z}=\gamma m\hbar .}

Spin behavior in a magnetic field

Consider nuclei which have a spin of one-half, like 1
H
, 13
C
or 19
F
. The nucleus has two possible spin states: m = 1/2 or m = -1/2 (also referred to as spin-up and spin-down, or sometimes ? and ? spin states, respectively). These states are degenerate; that is, they have the same energy. Hence the number of atoms in these two states will be approximately equal at thermal equilibrium.

If a nucleus is placed in a magnetic field, however, the interaction between the nuclear magnetic moment and the external magnetic field mean the two states no longer have the same energy. The energy of a magnetic moment ? when in a magnetic field B0 is given by:

E = - ? ? B 0 = - ? x B 0 x - ? y B 0 y - ? z B 0 z . {\displaystyle E=-{\boldsymbol {\mu }}\cdot \mathbf {B} _{0}=-\mu _{x}B_{0x}-\mu _{y}B_{0y}-\mu _{z}B_{0z}.}

Usually the z- axis is chosen to be along B0, and the above expression reduces to:

E = - ? z B 0   , {\displaystyle E=-\mu _{\mathrm {z} }B_{0}\ ,}

or alternatively:

E = - ? m ? B 0   . {\displaystyle E=-\gamma m\hbar B_{0}\ .}

As a result, the different nuclear spin states have different energies in a non-zero magnetic field. In less formal language, we can talk about the two spin states of a spin 1/2 as being aligned either with or against the magnetic field. If ? is positive (true for most isotopes) then m = 1/2 is the lower energy state.

The energy difference between the two states is:

? E = ? ? B 0   , {\displaystyle \Delta {E}=\gamma \hbar B_{0}\ ,}

and this difference results in a small population bias toward the lower energy state.

Magnetic resonance by nuclei

Resonant absorption by nuclear spins will occur only when electromagnetic radiation of the correct frequency (e.g., equaling the Larmor precession rate) is being applied to match the energy difference between the nuclear spin levels in a constant magnetic field of the appropriate strength. The energy of an absorbed photon is then E = h?0, where ?0 is the resonance radiofrequency that has to match (that is, it has to be equal to the Larmor precession frequency ?L of the nuclear magnetization in the constant magnetic field B0). Hence, a magnetic resonance absorption will only occur when ?E = h?0, which is when ?0 = ?B0/2?. Such magnetic resonance frequencies typically correspond to the radio frequency (or RF) range of the electromagnetic spectrum for magnetic fields up to roughly 20 T. It is this magnetic resonant absorption which is detected in NMR.

Nuclear shielding

It might appear from the above that all nuclei of the same nuclide (and hence the same ?) would resonate at the same frequency. This is not the case. The most important perturbation of the NMR frequency for applications of NMR is the "shielding" effect of the surrounding shells of electrons. Electrons, similar to the nucleus, are also charged and rotate with a spin to produce a magnetic field opposite to the magnetic field produced by the nucleus. In general, this electronic shielding reduces the magnetic field at the nucleus (which is what determines the NMR frequency).

As a result, the energy gap is reduced, and the frequency required to achieve resonance is also reduced. This shift in the NMR frequency due to the electronic molecular orbital coupling to the external magnetic field is called chemical shift, and it explains why NMR is able to probe the chemical structure of molecules, which depends on the electron density distribution in the corresponding molecular orbitals. If a nucleus in a specific chemical group is shielded to a higher degree by a higher electron density of its surrounding molecular orbital, then its NMR frequency will be shifted "upfield" (that is, a lower chemical shift), whereas if it is less shielded by such surrounding electron density, then its NMR frequency will be shifted "downfield" (that is, a higher chemical shift).

Unless the local symmetry of such molecular orbitals is very high (leading to "isotropic" shift), the shielding effect will depend on the orientation of the molecule with respect to the external field (B0). In solid-state NMR spectroscopy, magic angle spinning is required to average out this orientation dependence in order to obtain values close to the average chemical shifts. This is unnecessary in conventional NMR investigations of molecules, since rapid "molecular tumbling" averages out the chemical shift anisotropy (CSA). In this case, the term "average" chemical shift (ACS) is used.

Relaxation

The process called population relaxation refers to nuclei that return to the thermodynamic state in the magnet. This process is also called T1, "spin-lattice" or "longitudinal magnetic" relaxation, where T1 refers to the mean time for an individual nucleus to return to its thermal equilibrium state of the spins. Once the nuclear spin population is relaxed, it can be probed again, since it is in the initial, equilibrium (mixed) state.

The precessing nuclei can also fall out of alignment with each other (returning the net magnetization vector to a non-precessing field) and stop producing a signal. This is called T2 or transverse relaxation. Because of the difference in the actual relaxation mechanisms involved (for example, intermolecular versus intramolecular magnetic dipole-dipole interactions ), T1 is usually (except in rare cases) longer than T2 (that is, slower spin-lattice relaxation, for example because of smaller dipole-dipole interaction effects). In practice, the value of T2* which is the actually observed decay time of the observed NMR signal, or free induction decay (to 1/e of the initial amplitude immediately after the resonant RF pulse), also depends on the static magnetic field inhomogeneity, which is quite significant. (There is also a smaller but significant contribution to the observed FID shortening from the RF inhomogeneity of the resonant pulse). In the corresponding FT-NMR spectrum--meaning the Fourier transform of the free induction decay--the T2* time is inversely related to the width of the NMR signal in frequency units. Thus, a nucleus with a long T2 relaxation time gives rise to a very sharp NMR peak in the FT-NMR spectrum for a very homogeneous ("well-shimmed") static magnetic field, whereas nuclei with shorter T2 values give rise to broad FT-NMR peaks even when the magnet is shimmed well. Both T1 and T2 depend on the rate of molecular motions as well as the gyromagnetic ratios of both the resonating and their strongly interacting, next-neighbor nuclei that are not at resonance.

A Hahn echo decay experiment can be used to measure the dephasing time, as shown in the animation below. The size of the echo is recorded for different spacings of the two pulses. This reveals the decoherence which is not refocused by the 180° pulse. In simple cases, an exponential decay is measured which is described by the T2 time.


Simple explanation of the Proton NMR Spectrometer. - YouTube
src: i.ytimg.com


NMR spectroscopy

NMR spectroscopy is one of the principal techniques used to obtain physical, chemical, electronic and structural information about molecules due to either the chemical shift, Zeeman effect, or the Knight shift effect, or a combination of both, on the resonant frequencies of the nuclei present in the sample. It is a powerful technique that can provide detailed information on the topology, dynamics and three-dimensional structure of molecules in solution and the solid state. Thus, structural and dynamic information is obtainable (with or without "magic angle" spinning (MAS)) from NMR studies of quadrupolar nuclei (that is, those nuclei with spin S > 1/2) even in the presence of magnetic "dipole-dipole" interaction broadening (or simply, dipolar broadening) which is always much smaller than the quadrupolar interaction strength because it is a magnetic vs. an electric interaction effect.

Additional structural and chemical information may be obtained by performing double-quantum NMR experiments for quadrupolar nuclei such as 2
H
. Also, nuclear magnetic resonance is one of the techniques that has been used to design quantum automata, and also build elementary quantum computers.

Continuous-wave (CW) spectroscopy

In its first few decades, nuclear magnetic resonance spectrometers used a technique known as continuous-wave spectroscopy (CW spectroscopy). Although NMR spectra could be, and have been, obtained using a fixed magnetic field and sweeping the frequency of the electromagnetic radiation, this more typically involved using a fixed frequency source and varying the current (and hence magnetic field) in an electromagnet to observe the resonant absorption signals. This is the origin of the counterintuitive, but still common, "high field" and "low field" terminology for low frequency and high frequency regions respectively of the NMR spectrum.

As of 1996, CW instruments were still used for routine work because the older instruments were cheaper to maintain and operate, often operating at 60 MHz with correspondingly weaker (non-superconducting) electromagnets cooled with water rather than liquid helium. One radio coil operated continuously, sweeping through a range of frequencies, while another orthogonal coil, designed not to receive radiation from the transmitter, received signals from nuclei that reoriented in solution. As of 2014, low-end refurbished 60 MHz and 90 MHz systems were sold as FT-NMR instruments, and in 2010 the "average workhorse" NMR instrument was configured for 300 MHz.

CW spectroscopy is inefficient in comparison with Fourier analysis techniques (see below) since it probes the NMR response at individual frequencies in succession. Since the NMR signal is intrinsically weak, the observed spectrum suffers from a poor signal-to-noise ratio. This can be mitigated by signal averaging i.e. adding the spectra from repeated measurements. While the NMR signal is constant between scans and so adds linearly, the random noise adds more slowly - proportional to the square root of the number of spectra (see random walk). Hence the overall signal-to-noise ratio increases as the square-root of the number of spectra measured.

Fourier-transform spectroscopy

Most applications of NMR involve full NMR spectra, that is, the intensity of the NMR signal as a function of frequency. Early attempts to acquire the NMR spectrum more efficiently than simple CW methods involved illuminating the target simultaneously with more than one frequency. A revolution in NMR occurred when short pulses of radio-frequency radiation began to be used--centered at the middle of the NMR spectrum. In simple terms, a short pulse of a given "carrier" frequency "contains" a range of frequencies centered about the carrier frequency, with the range of excitation (bandwidth) being inversely proportional to the pulse duration, i.e. the Fourier transform of a short pulse contains contributions from all the frequencies in the neighborhood of the principal frequency. The restricted range of the NMR frequencies made it relatively easy to use short (millisecond to microsecond) radio frequency pulses to excite the entire NMR spectrum.

Applying such a pulse to a set of nuclear spins simultaneously excites all the single-quantum NMR transitions. In terms of the net magnetization vector, this corresponds to tilting the magnetization vector away from its equilibrium position (aligned along the external magnetic field). The out-of-equilibrium magnetization vector precesses about the external magnetic field vector at the NMR frequency of the spins. This oscillating magnetization vector induces a current in a nearby pickup coil, creating an electrical signal oscillating at the NMR frequency. This signal is known as the free induction decay (FID), and it contains the vector sum of the NMR responses from all the excited spins. In order to obtain the frequency-domain NMR spectrum (NMR absorption intensity vs. NMR frequency) this time-domain signal (intensity vs. time) must be Fourier transformed. Fortunately the development of Fourier Transform NMR coincided with the development of digital computers and the digital Fast Fourier Transform. Fourier methods can be applied to many types of spectroscopy. (See the full article on Fourier transform spectroscopy.)

Richard R. Ernst was one of the pioneers of pulse NMR, and he won a Nobel Prize in chemistry in 1991 for his work on Fourier Transform NMR and his development of multi-dimensional NMR (see below).

Multi-dimensional NMR spectroscopy

The use of pulses of different shapes, frequencies and durations in specifically designed patterns or pulse sequences allows the spectroscopist to extract many different types of information about the molecule. Multi-dimensional nuclear magnetic resonance spectroscopy is a kind of FT NMR in which there are at least two pulses and, as the experiment is repeated, the pulse sequence is systematically varied. In multidimensional nuclear magnetic resonance there will be a sequence of pulses and, at least, one variable time period. In three dimensions, two time sequences will be varied. In four dimensions, three will be varied.

There are many such experiments. In one, these time intervals allow (amongst other things) magnetization transfer between nuclei and, therefore, the detection of the kinds of nuclear-nuclear interactions that allowed for the magnetization transfer. Interactions that can be detected are usually classified into two kinds. There are through-bond interactions and through-space interactions, the latter usually being a consequence of the nuclear Overhauser effect. Experiments of the nuclear Overhauser variety may be employed to establish distances between atoms, as for example by 2D-FT NMR of molecules in solution.

Although the fundamental concept of 2D-FT NMR was proposed by Jean Jeener from the Free University of Brussels at an International Conference, this idea was largely developed by Richard Ernst who won the 1991 Nobel prize in Chemistry for his work in FT NMR, including multi-dimensional FT NMR, and especially 2D-FT NMR of small molecules. Multi-dimensional FT NMR experiments were then further developed into powerful methodologies for studying biomolecules in solution, in particular for the determination of the structure of biopolymers such as proteins or even small nucleic acids.

In 2002 Kurt Wüthrich shared the Nobel Prize in Chemistry (with John Bennett Fenn and Koichi Tanaka) for his work with protein FT NMR in solution.

Solid-state NMR spectroscopy

This technique complements X-ray crystallography in that it is frequently applicable to molecules in a liquid or liquid crystal phase, whereas crystallography, as the name implies, is performed on molecules in a solid phase. Though nuclear magnetic resonance is used to study solids, extensive atomic-level molecular structural detail is especially challenging to obtain in the solid state. There is little signal averaging by thermal motion in the solid state, where most molecules can only undergo restricted vibrations and rotations at room temperature, each in a slightly different electronic environment, therefore exhibiting a different NMR absorption peak. Such a variation in the electronic environment of the resonating nuclei results in a blurring of the observed spectra--which is often only a broad Gaussian band for non-quadrupolar spins in a solid- thus making the interpretation of such "dipolar" and "chemical shift anisotropy" (CSA) broadened spectra either very difficult or impossible.

Professor Raymond Andrew at the University of Nottingham in the UK pioneered the development of high-resolution solid-state nuclear magnetic resonance. He was the first to report the introduction of the MAS (magic angle sample spinning; MASS) technique that allowed him to achieve spectral resolution in solids sufficient to distinguish between chemical groups with either different chemical shifts or distinct Knight shifts. In MASS, the sample is spun at several kilohertz around an axis that makes the so-called magic angle ?m (which is ~54.74°, where cos2 ?m = 1/3) with respect to the direction of the static magnetic field B0; as a result of such magic angle sample spinning, the chemical shift anisotropy bands are averaged to their corresponding average (isotropic) chemical shift values. The above expression involving cos2 ?m has its origin in a calculation that predicts the magnetic dipolar interaction effects to cancel out for the specific value of ?m called the magic angle. One notes that correct alignment of the sample rotation axis as close as possible to ?m is essential for cancelling out the dipolar interactions whose strength for angles sufficiently far from ?m is usually greater than ~10 kHz for C-H bonds in solids, for example, and it is thus greater than their CSA values.

There are different angles for the sample spinning relative to the applied field for the averaging of quadrupole interactions and paramagnetic interactions, correspondingly ~30.6° and ~70.1°.

A concept developed by Sven Hartmann and Erwin Hahn was utilized in transferring magnetization from protons to less sensitive nuclei (popularly known as cross-polarization) by M.G. Gibby, Alex Pines and John S. Waugh. Then, Jake Schaefer and Ed Stejskal demonstrated also the powerful use of cross-polarization under MASS conditions (CP-MAS) which is now routinely employed to measure high resolution spectra of low-abundance and low-sensitivity nuclei, namely carbon-13, in solids.

Sensitivity

Because the intensity of nuclear magnetic resonance signals and, hence, the sensitivity of the technique depends on the strength of the magnetic field the technique has also advanced over the decades with the development of more powerful magnets. Advances made in audio-visual technology have also improved the signal-generation and processing capabilities of newer instruments.

As noted above, the sensitivity of nuclear magnetic resonance signals is also dependent on the presence of a magnetically susceptible nuclide and, therefore, either on the natural abundance of such nuclides or on the ability of the experimentalist to artificially enrich the molecules, under study, with such nuclides. The most abundant naturally occurring isotopes of hydrogen and phosphorus (for example) are both magnetically susceptible and readily useful for nuclear magnetic resonance spectroscopy. In contrast, carbon and nitrogen have useful isotopes but which occur only in very low natural abundance.

Other limitations on sensitivity arise from the quantum-mechanical nature of the phenomenon. For quantum states separated by energy equivalent to radio frequencies, thermal energy from the environment causes the populations of the states to be close to equal. Since incoming radiation is equally likely to cause stimulated emission (a transition from the upper to the lower state) as absorption, the NMR effect depends on an excess of nuclei in the lower states. Several factors can reduce sensitivity, including:

  • Increasing temperature, which evens out the population of states. Conversely, low temperature NMR can sometimes yield better results than room-temperature NMR, providing the sample remains liquid.
  • Saturation of the sample with energy applied at the resonant radiofrequency. This manifests in both CW and pulsed NMR; in the first case (CW) this happens by using too much continuous power that keeps the upper spin levels completely populated; in the second case (pulsed), each pulse (that is at least a 90° pulse) leaves the sample saturated, and four to five times the (longitudinal) relaxation time (5T1) must pass before the next pulse or pulse sequence can be applied. For single pulse experiments, shorter RF pulses that tip the magnetization by less than 90° can be used, which loses some intensity of the signal, but allows for shorter recycle delays. The optimum there is called an Ernst angle, after the Nobel laureate. Especially in solid state NMR, or in samples with very few nuclei with spins greater than zero, (diamond with the natural 1% of carbon-13 is especially troublesome here) the longitudinal relaxation times can be on the range of hours, while for proton-NMR they are more on the range of one second.
  • Non-magnetic effects, such as electric-quadrupole coupling of spin-1 and spin-3/2 nuclei with their local environment, which broaden and weaken absorption peaks. 14
    N
    , an abundant spin-1 nucleus, is difficult to study for this reason. High resolution NMR instead probes molecules using the rarer 15
    N
    isotope, which has spin-1/2.

Isotopes

Many isotopes of chemical elements can be used for NMR analysis.

Commonly used nuclei:

  • 1
    H
    , the most commonly used spin-1/2 nucleus in NMR investigation, has been studied using many forms of NMR. Hydrogen is highly abundant, especially in biological systems. It is the nucleus most sensitive to NMR signal (apart from 3
    H
    which is not commonly used due to its instability and radioactivity). Proton NMR produces narrow chemical shift with sharp signals. Fast acquisition of quantitative results (peak integrals in stoichiometric ratio) is possible due to short relaxation time. The 1
    H
    signal has been the sole diagnostic nucleus used for clinical magnetic resonance imaging.
  • 2
    H
    , a spin 1 nucleus commonly utilized as signal-free medium in the form of deuterated solvents during proton NMR, to avoid signal interference from hydrogen-containing solvents in measurement of 1
    H
    solutes. Also used in determining the behavior of lipids in lipid membranes and other solids or liquid crystals as it is a relatively non-perturbing label which can selectively replace 1
    H
    . Alternatively, 2
    H
    can be detected in media specially labeled with 2
    H
    . Deuterium resonance is commonly used in high-resolution NMR spectroscopy to monitor drifts in the magnetic field strength (lock) and to improve the homogeneity of the external magnetic field.
  • 3
    He
    , is very sensitive to NMR. There is a very low percentage in natural helium, and subsequently has to be purified from 4
    He
    . It is used mainly in studies of endohedral fullerenes, where its chemical inertness is beneficial to ascertaining the structure of the entrapping fullerene.
  • 11
    B
    , more sensitive than 10
    B
    , yields sharper signals. Quartz tubes must be used as borosilicate glass interferes with measurement.
  • 13
    C
    spin-1/2, is widely used, despite its relative paucity in naturally occurring carbon (approximately 1%). It is stable to nuclear decay. Since there is a low percentage in natural carbon, spectrum acquisition on samples which have not been experimentally enriched in 13
    C
    takes a long time. Frequently used for labeling of compounds in synthetic and metabolic studies. Has low sensitivity and wide chemical shift, yields sharp signals. Low percentage makes it useful by preventing spin-spin couplings and makes the spectrum appear less crowded. Slow relaxation means that spectra are not integrable unless long acquisition times are used.
  • 14
    N
    , spin-1, medium sensitivity nucleus with wide chemical shift. Its large quadrupole moment interferes in acquisition of high resolution spectra, limiting usefulness to smaller molecules and functional groups with a high degree of symmetry such as the headgroups of lipids.
  • 15
    N
    , spin-1/2, relatively commonly used. Can be used for labeling compounds. Nucleus very insensitive but yields sharp signals. Low percentage in natural nitrogen together with low sensitivity requires high concentrations or expensive isotope enrichment.
  • 17
    O
    , spin-5/2, low sensitivity and very low natural abundance (0.037%), wide chemical shifts range (up to 2000 ppm). Quadrupole moment causing a line broadening. Used in metabolic and biochemical studies in studies of chemical equilibria.
  • 19
    F
    , spin-1/2, relatively commonly measured. Sensitive, yields sharp signals, has wide chemical shift.
  • 31
    P
    , spin-1/2, 100% of natural phosphorus. Medium sensitivity, wide chemical shifts range, yields sharp lines. Spectra tend to have a moderate amount of noise. Used in biochemical studies and in coordination chemistry where phosphorus containing ligands are involved.
  • 35
    Cl
    and 37
    Cl
    , broad signal. 35
    Cl
    significantly more sensitive, preferred over 37
    Cl
    despite its slightly broader signal. Organic chlorides yield very broad signals, its use is limited to inorganic and ionic chlorides and very small organic molecules.
  • 43
    Ca
    , used in biochemistry to study calcium binding to DNA, proteins, etc. Moderately sensitive, very low natural abundance.
  • 195
    Pt
    , used in studies of catalysts and complexes.

Other nuclei (usually used in the studies of their complexes and chemical binding, or to detect presence of the element):


LEGO Ideas - NMR Spectrometer
src: ideascdn.lego.com


Applications

NMR is extensively used in medicine in the form of magnetic resonance imaging. NMR is used industrially mainly for routine analysis of chemicals. The technique is also used, for example, to measure the ratio between water and fat in foods, monitoring of flow of corrosive fluids in pipes, or to study molecular structures such as catalysts.

Medicine

The application of nuclear magnetic resonance best known to the general public is magnetic resonance imaging for medical diagnosis and magnetic resonance microscopy in research settings, however, it is also widely used in chemical studies, notably in NMR spectroscopy such as proton NMR, carbon-13 NMR, deuterium NMR and phosphorus-31 NMR. Biochemical information can also be obtained from living tissue (e.g. human brain tumors) with the technique known as in vivo magnetic resonance spectroscopy or chemical shift NMR Microscopy.

These studies are possible because nuclei are surrounded by orbiting electrons, which are charged particles that generate small, local magnetic fields that add to or subtract from the external magnetic field, and so will partially shield the nuclei. The amount of shielding depends on the exact local environment. For example, a hydrogen bonded to an oxygen will be shielded differently from a hydrogen bonded to a carbon atom. In addition, two hydrogen nuclei can interact via a process known as spin-spin coupling, if they are on the same molecule, which will split the lines of the spectra in a recognizable way.

As one of the two major spectroscopic techniques used in metabolomics, NMR is used to generate metabolic fingerprints from biological fluids to obtain information about disease states or toxic insults.

Chemistry

By studying the peaks of nuclear magnetic resonance spectra, chemists can determine the structure of many compounds. It can be a very selective technique, distinguishing among many atoms within a molecule or collection of molecules of the same type but which differ only in terms of their local chemical environment. NMR spectroscopy is used to unambiguously identify known and novel compounds, and as such, is usually required by scientific journals for identity confirmation of synthesized new compounds. See the articles on carbon-13 NMR and proton NMR for detailed discussions.

By studying T2 information, a chemist can determine the identity of a compound by comparing the observed nuclear precession frequencies to known frequencies. Further structural data can be elucidated by observing spin-spin coupling, a process by which the precession frequency of a nucleus can be influenced by the magnetization transfer from nearby chemically bound nuclei. Spin-spin coupling is observed in NMR of hydrogen-1 (1
H
 NMR), since its natural abundance is nearly 100%; isotope enrichment is required for most other elements.

Because the nuclear magnetic resonance timescale is rather slow, compared to other spectroscopic methods, changing the temperature of a T2* experiment can also give information about fast reactions, such as the Cope rearrangement or about structural dynamics, such as ring-flipping in cyclohexane. At low enough temperatures, a distinction can be made between the axial and equatorial hydrogens in cyclohexane.

An example of nuclear magnetic resonance being used in the determination of a structure is that of buckminsterfullerene (often called "buckyballs", composition C60). This now famous form of carbon has 60 carbon atoms forming a sphere. The carbon atoms are all in identical environments and so should see the same internal H field. Unfortunately, buckminsterfullerene contains no hydrogen and so 13
C
nuclear magnetic resonance has to be used. 13
C
spectra require longer acquisition times since carbon-13 is not the common isotope of carbon (unlike hydrogen, where 1
H
is the common isotope). However, in 1990 the spectrum was obtained by R. Taylor and co-workers at the University of Sussex and was found to contain a single peak, confirming the unusual structure of buckminsterfullerene.

Purity determination (w/w NMR)

NMR is primarily used for structural determination, however it can also be used for purity determination, providing that the structure and molecular weight of the compound is known. This technique requires the use of an internal standard of a known purity. Typically this standard will have a high molecular weight to facilitate accurate weighing, but relatively few protons so as to give a clear peak for later integration e.g. 1,2,4,5-tetrachloro-3-nitrobenzene. Accurately weighed portions of the standard and sample are combined and analysed by NMR. Suitable peaks are selected for both compounds and the purity of the sample determined via the following equation.

P u r i t y = w s t d × n [ H ] s t d × M W s p l w s p l × M W s t d × n [ H ] s p l × P {\displaystyle \mathrm {Purity} ={\frac {w_{\mathrm {std} }\times n[\mathrm {H} ]_{\mathrm {std} }\times MW_{\mathrm {spl} }}{w_{\mathrm {spl} }\times MW_{\mathrm {std} }\times n[\mathrm {H} ]_{\mathrm {spl} }}}\times P}

Where:

  • wstd: weight of internal standard
  • wspl: weight of sample
  • n[H]std: the integrated area of the peak selected for comparison in the standard, corrected for the number of protons in that functional group
  • n[H]spl: the integrated area of the peak selected for comparison in the sample, corrected for the number of protons in that functional group
  • MWstd: molecular weight of standard
  • MWspl: molecular weight of sample
  • P: purity of internal standard

Non-destructive testing

Nuclear magnetic resonance is extremely useful for analyzing samples non-destructively. Radio waves and static magnetic fields easily penetrate many types of matter and anything that is not inherently ferromagnetic. For example, various expensive biological samples, such as nucleic acids, including RNA and DNA, or proteins, can be studied using nuclear magnetic resonance for weeks or months before using destructive biochemical experiments. This also makes nuclear magnetic resonance a good choice for analyzing dangerous samples.

Acquisition of dynamic information

In addition to providing static information on molecules by determining their 3D structures in solution, one of the remarkable advantages of NMR over X-ray crystallography is that it can be used to obtain important dynamic information.

Data acquisition in the petroleum industry

Another use for nuclear magnetic resonance is data acquisition in the petroleum industry for petroleum and natural gas exploration and recovery. Initial research in this domain began in the 1950s, however, the first commercial instruments were not released until the early 1990s. A borehole is drilled into rock and sedimentary strata into which nuclear magnetic resonance logging equipment is lowered. Nuclear magnetic resonance analysis of these boreholes is used to measure rock porosity, estimate permeability from pore size distribution and identify pore fluids (water, oil and gas). These instruments are typically low field NMR spectrometers.

NMR logging, a subcategory of electromagnetic logging, measures the induced magnet moment of hydrogen nuclei (protons) contained within the fluid-filled pore space of porous media (reservoir rocks). Unlike conventional logging measurements (e.g., acoustic, density, neutron, and resistivity), which respond to both the rock matrix and fluid properties and are strongly dependent on mineralogy, NMR-logging measurements respond to the presence of hydrogen protons. Because these protons primarily occur in pore fluids, NMR effectively responds to the volume, composition, viscosity, and distribution of these fluids, for example oil, gas or water. NMR logs provide information about the quantities of fluids present, the properties of these fluids, and the sizes of the pores containing these fluids. From this information, it is possible to infer or estimate:

  • The volume (porosity) and distribution (permeability) of the rock pore space
  • Rock composition
  • Type and quantity of fluid hydrocarbons
  • Hydrocarbon producibility

The basic core and log measurement is the T2 decay, presented as a distribution of T2 amplitudes versus time at each sample depth, typically from 0.3 ms to 3 s. The T2 decay is further processed to give the total pore volume (the total porosity) and pore volumes within different ranges of T2. The most common volumes are the bound fluid and free fluid. A permeability estimate is made using a transform such as the Timur-Coates or SDR permeability transforms. By running the log with different acquisition parameters, direct hydrocarbon typing and enhanced diffusion are possible.

Flow probes for NMR spectroscopy

Recently, real-time applications of NMR in liquid media have been developed using specifically designed flow probes (flow cell assemblies) which can replace standard tube probes. This has enabled techniques that can incorporate the use of high performance liquid chromatography (HPLC) or other continuous flow sample introduction devices.

Process control

NMR has now entered the arena of real-time process control and process optimization in oil refineries and petrochemical plants. Two different types of NMR analysis are utilized to provide real time analysis of feeds and products in order to control and optimize unit operations. Time-domain NMR (TD-NMR) spectrometers operating at low field (2-20 MHz for 1
H
) yield free induction decay data that can be used to determine absolute hydrogen content values, rheological information, and component composition. These spectrometers are used in mining, polymer production, cosmetics and food manufacturing as well as coal analysis. High resolution FT-NMR spectrometers operating in the 60 MHz range with shielded permanent magnet systems yield high resolution 1
H
NMR spectra of refinery and petrochemical streams. The variation observed in these spectra with changing physical and chemical properties is modeled using chemometrics to yield predictions on unknown samples. The prediction results are provided to control systems via analogue or digital outputs from the spectrometer.

Earth's field NMR

In the Earth's magnetic field, NMR frequencies are in the audio frequency range, or the very low frequency and ultra low frequency bands of the radio frequency spectrum. Earth's field NMR (EFNMR) is typically stimulated by applying a relatively strong dc magnetic field pulse to the sample and, after the end of the pulse, analyzing the resulting low frequency alternating magnetic field that occurs in the Earth's magnetic field due to free induction decay (FID). These effects are exploited in some types of magnetometers, EFNMR spectrometers, and MRI imagers. Their inexpensive portable nature makes these instruments valuable for field use and for teaching the principles of NMR and MRI.

An important feature of EFNMR spectrometry compared with high-field NMR is that some aspects of molecular structure can be observed more clearly at low fields and low frequencies, whereas other aspects observable at high fields are not observable at low fields. This is because:

  • Electron-mediated heteronuclear J-couplings (spin-spin couplings) are field independent, producing clusters of two or more frequencies separated by several Hz, which are more easily observed in a fundamental resonance of about 2 kHz."Indeed it appears that enhanced resolution is possible due to the long spin relaxation times and high field homogeneity which prevail in EFNMR."
  • Chemical shifts of several ppm are clearly separated in high field NMR spectra, but have separations of only a few millihertz at proton EFNMR frequencies, so are usually lost in noise etc.

Zero field NMR

In zero field NMR all magnetic fields are shielded such that magnetic fields below 1 nT (nanotesla) are achieved and the nuclear precession frequencies of all nuclei are close to zero and indistinguishable. Under those circumstances the observed spectra are no-longer dictated by chemical shifts but primarily by J-coupling interactions which are independent of the external magnetic field. Since inductive detection schemes are not sensitive at very low frequencies, on the order of the J-couplings (typically between 0 and 1000 Hz), alternative detection schemes are used. Specifically, sensitive magnetometers turn out to be good detectors for zero field NMR. A zero magnetic field environment does not provide any polarization hence it is the combination of zero field NMR with hyperpolarization schemes that makes zero field NMR attractive.

Quantum computing

NMR quantum computing uses the spin states of nuclei within molecules as qubits. NMR differs from other implementations of quantum computers in that it uses an ensemble of systems, in this case molecules.

Magnetometers

Various magnetometers use NMR effects to measure magnetic fields, including proton precession magnetometers (PPM) (also known as proton magnetometers), and Overhauser magnetometers. See also Earth's field NMR.


Nuclear Magnetic Resonance With No Magnets | Berkeley Lab
src: newscenter.lbl.gov


Makers of NMR equipment

Major NMR instrument makers include Thermo Fisher Scientific, Magritek, Oxford Instruments, Bruker, Spinlock SRL, General Electric, JEOL, Kimble Chase, Philips, Siemens AG, and Agilent Technologies, Inc. (who own Varian, Inc.).


Isometric Flat 3D Concept Vector Hospital Medical Mri Web ...
src: previews.123rf.com


See also


Quantitative Serum Nuclear Magnetic Resonance Metabolomics in ...
src: circgenetics.ahajournals.org


References


Media Portfolio
src: wps.prenhall.com


Further reading

  • John D. Roberts (1959). Nuclear Magnetic Resonance : applications to organic chemistry. McGraw-Hill Book Company. ISBN 9781258811662. 
  • J.A.Pople; W.G.Schneider; H.J.Bernstein (1959). High-resolution Nuclear Magnetic Resonance. McGraw-Hill Book Company. 
  • A. Abragam (1961). The Principles of Nuclear Magnetism. Clarendon Press. ISBN 9780198520146. 
  • Charles P. Slichter (1963). Principles of magnetic resonance: with examples from solid state physics. Harper & Row. ISBN 9783540084761. 
  • John Emsley; James Feeney; Leslie Howard Sutcliffe (1965). High Resolution Nuclear Magnetic Resonance Spectroscopy. Pergamon. ISBN 9781483184081. 
  • David M. Grant; Robin Kingsley Harris (2002). Encyclopedia of Nuclear Magnetic Resonance: Advances in NMR. John Wiley. ISBN 9780471490821. 
  • Gary E. Martin; A. S. Zektzer (1988). Two-Dimensional NMR Methods for Establishing Molecular Connectivity. New York: Wiley-VCH. p. 59. ISBN 0-471-18707-0. 
  • J.W. Akitt; B.E. Mann (2000). NMR and Chemistry. Cheltenham, UK: Stanley Thornes. pp. 273, 287. ISBN 0-7487-4344-8. 
  • G.M. Clore; A.M. Gronenborn (1991). "Structures of larger proteins in solution: three- and four-dimensional heteronuclear NMR spectroscopy". Science. 252 (5011): 1390-1399. doi:10.1126/science.2047852. PMID 2047852. 
  • J.P. Hornak. "The Basics of NMR". Retrieved 2009-02-23. 
  • J. Keeler (2005). Understanding NMR Spectroscopy. John Wiley & Sons. ISBN 0-470-01786-4. 
  • Kurt Wüthrich (1986). NMR of Proteins and Nucleic Acids. New York (NY), USA: Wiley-Interscience. ISBN 0-471-11917-2. 
  • J.M. Tyszka; S.E. Fraser; R.E. Jacobs (2005). "Magnetic resonance microscopy: recent advances and applications". Current Opinion in Biotechnology. 16 (1): 93-99. doi:10.1016/j.copbio.2004.11.004. PMID 15722021. 
  • R.L. Haner; P.A. Keifer (2009). Encyclopedia of Magnetic Resonance. John Wiley. doi:10.1002/9780470034590.emrstm1085. 

Nuclear Magnetic Resonance Spectroscopy Stock Photos & Nuclear ...
src: l7.alamy.com


External links

Tutorial

  • NMR/MRI tutorial
  • NMR Library NMR Concepts
  • NMR Course Notes

Animations and simulations

  • This animation shows a spin, the modification of spin with magnetic field and HF pulse, spin echo sequences, inversion recovery sequence, gradient echo sequence and relaxation of spin
  • A free interactive simulation of NMR principles

Video

  • introduction to NMR and MRI
  • Richard Ernst, NL - Developer of Multdimensional NMR techniques Freeview video provided by the Vega Science Trust.
  • 'An Interview with Kurt Wuthrich' Freeview video by the Vega Science Trust (Wüthrich was awarded a Nobel Prize in Chemistry in 2002 "for his development of nuclear magnetic resonance spectroscopy for determining the three-dimensional structure of biological macromolecules in solution").
Other
  • Off Magic Angle Spinning

Source of the article : Wikipedia